EasyManua.ls Logo

Agilent Technologies ENA Series

Agilent Technologies ENA Series
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 73
Performance Tests
E5062A Performance Test Record (for Option 175/275)
2. Performance Test
CW Frequency: 1.5 GHz (relative to 0 dBm reference)
CW Frequency: 3 GHz (relative to 0 dBm reference)
Trace Noise Test
Direction: S21 (Magnitude)
Direction: S21 (Phase)
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-5 0.75 0.03
-2.5 0.75 0.03
2.5 0.75 0.03
5 0.75 0.03
7.5 0.75 0.03
10 0.75 0.03
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-5 0.75 0.03
-2.5 0.75 0.03
2.5 0.75 0.03
5 0.75 0.03
7.5 0.75 0.03
10 0.75 0.03
Frequency range
[Hz]
IF bandwidth
[Hz]
Test limit [dB rms] Test result [dB rms]
300 k - 1 M 3 k 0.008
1 M - 100 M 3 k 0.005
100 M - 1.5 G 3 k 0.005
1.5 G - 3 G 3 k 0.005
Frequency range
[Hz]
IF bandwidth
[Hz]
Test limit [deg
rms]
Test result [deg
rms]
300 k - 1 M 3 k 0.05