Chapter 2 43
Performance Tests
E5061A Performance Test Record (for Option 150/250)
2. Performance Test
Direction: S12 (Options 250 only)
Dynamic Accuracy Test
@ Reference power level: -10 dBm
Direction: S21
100 M - 1.5 G 10 -120
300 k - 1 M 3 k -90
1 M - 100 M 3 k -95
100 M - 1.5 G 3 k -95
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 1 M 10 -115
1 M - 100 M 10 -120
100 M - 1.5 G 10 -120
300 k - 1 M 3 k -90
1 M - 100 M 3 k -95
100 M - 1.5 G 3 k -95
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 0.303 0.011
5 0.144 0.011
0 0.090 0.011
-5 0.068 0.011
-15 0.065 0.011
-20 0.073 0.011
-25 0.080 0.015
-30 0.087 0.015
-35 0.095 0.019
-40 0.103 0.019
-45 0.111 0.022
-50 0.121 0.022

Table of Contents