EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G1364B

Agilent Technologies G1364B User Manual

Agilent Technologies G1364B
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
Agilent 1100 Series Fraction Collectors User’s Guide 33
Configuration and Operation of the Fraction Collector 1
Setting up a Fraction Collector Method
chromatogram can be zoomed, the values for up and down slope can be
specified and the upper and lower threshold level can be set-up. The
graphically specified values are automatically transferred to the Peak Detector
Table.
NOTE
The run time for a system with a fraction collector must be extended by the delay time, to
ensure the complete collection of all compounds.
The run time is calculated as:
Run time = end time of last collected peak + delay time (with delay time =
delay volume/flow rate)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1364B and is the answer not in the manual?

Agilent Technologies G1364B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1364B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals