Agilent 1100 Series Fraction Collectors User’s Guide 33
Configuration and Operation of the Fraction Collector 1
Setting up a Fraction Collector Method
chromatogram can be zoomed, the values for up and down slope can be
specified and the upper and lower threshold level can be set-up. The
graphically specified values are automatically transferred to the Peak Detector
Table.
NOTE
The run time for a system with a fraction collector must be extended by the delay time, to
ensure the complete collection of all compounds.
The run time is calculated as:
Run time = end time of last collected peak + delay time (with delay time =
delay volume/flow rate)