Agilent 1100 Series Fraction Collectors User’s Guide 59
Repairing the Fraction Collector 3
Simple Repairs
Replacing the Inlet/Waste Tubings
Frequency When contaminated, worn out or visibly damaged
Typically once every year
Tools requi re d None
Parts Required Inlet / waste tubing assembly included in tubing kit preparative scale 0.8 mm
ID, PN G1364-68711 or
Inlet / waste tubing assembly included in tubing kit
analytical scale 0.25 mm ID, PN G1364-68712
WARNING
To avoid personal injury, keep fingers away from the needle area during fraction
collector operation.
WARNING
Follow the described installation procedure exactly to maximize the lifetime of the
inlet / waste tubings and to avoid potential spills or fraction losses. Regularly
inspect the tubings and exchange them if they are worn out or show visible signs of
damage.