42 Agilent 1100 Series Fraction Collectors User’s Guide
1 Configuration and Operation of the Fraction Collector
Optimizing fraction collection
Optimizing fraction collection
Time-based
fraction
collection
Time slices must have a length of at least 0.05 min.
Set # of Fractions such that length of resulting fractions is at least 0.05 min.
Peak-based
fraction
collection
Set threshold and slope values such that length of fractions is at least 0.05
min.
Unresolved peaks can be separated using appropriate threshold and slope
values. If two unresolved peaks are to be collected as one fraction, collect
based on threshold only.
If the baseline of the chromatogram is below or above 0 mAU, this offset is
not accounted for when triggering peaks using a threshold value. The
threshold value is always added to 0 mAU.