Agilent 1100 Series Fraction Collectors User’s Guide 41
Configuration and Operation of the Fraction Collector 1
Special Applications
Using high test tubes in the analytical scale fraction collector
In the standard configuration the analytical scale fraction collector G1364C
the maximum test tube height is limited to 48 mm. This limitation can be
extended to 75 mm if
the standard 50 mm needle assembly (G1367-87200) is replaced by the
20 mm needle assembly (G1364-87202)
the wellplate adapter is removed.
Figure 19 Fraction collector configuration parameters for semi-prep operation