52 Agilent 1100 Series Fraction Collectors User’s Guide
2 Troubleshooting and Test Functions
Transport Unit Self Alignment
3 Ensure that the wellplate adapter is correctly
assembled
4 Select the Maintenance menu in the
Diagnosis view of the Agilent ChemStation.
5 In the menu choose Fraction Collector >
Transport Alignment... to start the automated
procedure.
The Transport Alignment Procedure takes
approximately 10-15 minutes
6 Set the 8-bit configuration switch to the default
setting.
See “8-bit configuration switch” in the
Fraction Collectors Service Manual.
How to perform a Transport Unit Self Alignment?
Steps Comments
NOTE
If the Transport Unit Self Alignment is started with well plates on the tray, the alignment
procedure is aborted without any error message.