EasyManuals Logo

Agilent Technologies G7129B User Manual

Agilent Technologies G7129B
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #239 background imageLoading...
Page #239 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 239
8Maintenance
Exchange the Needle Seat Assembly
13 Reconnect the sample loop to the needle. Make sure of
the correct positioning of the loop capillary, the
uncoated part of the capillary must be horizontal.
NOTE
Do not overtighten the fitting!
NOTE
Incorrect positioning and installation of the needle/loop
connection can result in damaging or breaking the sample
loop.
14 In Lab Advisor use Next to lift the needle slowly into the
up position.
OR
In the Local Controller, move the needle up to the upper-
most position.
15 Move the wash port back into its normal position by
turning it to the right.
16 Install the safety cover (1.) and fix the screw (2.).
(1.)
(2.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129B and is the answer not in the manual?

Agilent Technologies G7129B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals