Agilent InfinityLab LC Series Vialsamplers User Manual 240
8Maintenance
Exchange the Needle Seat Assembly
17 In Lab Advisor, select Next to move the needle into the
needle seat and then Back to finish the maintenance
procedure.
OR
In the Local Controller, select Continue to move the nee-
dle into the needle seat and then Exit to finish the main-
tenance procedure.
18 Close the doors.
19 Perform a System Pressure Test.