M3860A and M3861A HEARTSTREAM FR2 Defibrillator 2-1
2 Preparing Your HEARTSTREAM FR2 for Use
Overview
There are a few basic steps to preparing your HEARTSTREAM FR2 for use:
Install a battery.
Set the clock in the FR2 (optional).
Run the battery insertion selftest.
Place the FR2 with recommended accessories in a convenient location.
The instructions presented here briefly describe the normal sequence of
preparation. It assumes that you are using a fresh battery, that the selftest
passes, and that the factory default settings will not be changed. Exceptions to
this sequence are provided elsewhere in this User’s Guide.
Installing the Battery
The HEARTSTREAM FR2 is shipped with a M3863A battery pack. The battery
is enclosed in a gray plastic case. There is a yellow latch at one end that holds
the battery pack in place when it is correctly installed in the FR2. To install
the battery:
1. Hold the battery pack by the latch end and slide it into the battery
compartment at the top of the H
EARTSTREAM FR2.
2. Slide the battery all the way into the opening, until the latch clicks into
place. The latch will click into place only when the battery is inserted
correctly.
CAUTION: Follow all instructions supplied with the HEARTSTREAM M3863A battery pack. Install the
battery before the expiration date shown on the battery.