M3860A and M3861A HEARTSTREAM FR2 Defibrillator B-1
B Technical Specifications
The specifications for the HEARTSTREAM FR2 provided in this chapter apply
to both the M3860A and M3861A, unless otherwise noted.
HEARTSTREAM FR2 AED Specifications
Physical
Environmental
CATEGORY NOMINAL SPECIFICATIONS
Size 2.6" high x 8.6" wide x 8.6" deep (6.6 cm x 21.8 cm x 21.8 cm).
Weight Approximately 4.7 lbs (2 kg) with battery installed.
CATEGORY NOMINAL SPECIFICATIONS
Operating Temperature
and Humidity
32° to 122° F (0° to 50° C)
0% to 95% relative humidity (non-condensing).
Standby Temperature
and Humidity
32° to 109° F (0° to 43° C).
0% to 75% relative humidity (non-condensing).
Applies to H
EARTSTREAM FR2 with battery installed and stored with defibrillation
pads.
Altitude Meets MIL-810E 500.3, Procedure II (-500 feet to 15,000 feet).
Shock/Drop Abuse
To l e r a n c e
Meets MIL-STD-810E 516.4, Procedure IV (after a 1 meter drop to any edge,
corner, or surface, in standby mode).
Vibration Meets MIL-STD-810E 514.4-17.