Setup Overview
6-4 Setup and Advanced Mode Features
Manual Override Parameters
The HEARTSTREAM FR2 provides several ways of overriding the automatic
protocol. The parameters in the following table are used to enable different
kinds of manual override.
CPR PROMPT LONG, SHORT LONG Sets the level of detail provided in the CPR reminder voice
prompts provided at the completion of a Shock Series.
LONG — provides detailed coaching to check airway,
breath, and pulse before beginning CPR.
SHORT — simply directs user to begin CPR if needed.
MONITOR
PROMPT
INTERVAL
(minutes)
0.5, 1.0, 1.5,
2.0, 2.5, 3.0,
=
(infinite)
1.0 Sets the interval for patient care prompts provided during
FR2 monitoring of the patient’s ECG following an NSA
decision. Selection of
(infinite) means no repeat
prompting will be provided during ECG monitoring.
PAR AM ETE R SETTINGS DEFAULT DESCRIPTION
PARAM ET ER SETTINGS DEFAULT DESCRIPTION
ADVANCED OFF,
ANALYZE,
CHARGE
OFF Enables or disables advanced mode entry for ALS or
tiered-response systems.
OFF — disables advanced mode features.
ANALYZE — enables user-initiated rhythm analysis and
disarm, and (M3860A only) turns on ECG display.
CHARGE (M3860A only) — in addition to enabling the
analyze feature, enables user-initiated charging and
disarming.