Maintenance
M3860A and M3861A H
EARTSTREAM FR2 Defibrillator 4-13
4
Battery History
Information about use of the battery currently installed in your HEARTSTREAM
FR2 is also available. To review the history of the battery, select NEXT from
the menu screen displayed when you insert the battery, then select BATTERY
HISTORY from the next menu displayed.
The battery history information is automatically stored in the internal memory
of the battery. It includes:
USE MINUTES — the total operating time (in minutes), including
selftest time, for this battery;
CHARGES — the total number of full defibrillation charges that have
been provided by this battery;
BATTERY — a GOOD BATTERY, LOW BATTERY or REPLACE
BATTERY message, as appropriate, and
STATUS — the current status of this battery, displayed in a binary code
that should be recorded if technical service is needed.
Maintenance
Maintenance Schedule
Maintenance of the HEARTSTREAM FR2 is very simple, but it is a very
important factor in its dependability. The H
EARTSTREAM FR2 performs many
maintenance activities itself. These include daily and weekly selftests to
verify readiness for use and more detailed monthly selftests that also verify
the shock waveform delivery system. In addition, a detailed selftest is run
whenever a battery is installed in the FR2.
The H
EARTSTREAM FR2 requires no calibration or verification of energy
delivery. The H
EARTSTREAM FR2 has no user-serviceable parts.
CAUTION: Improper maintenance may damage the HEARTSTREAM FR2 or cause it to function
improperly. Maintain the H
EARTSTREAM FR2 only as described in this User's Guide or as designated by
your program's Medical Director.