Using Advanced Mode Features
6-12 Setup and Advanced Mode Features
Press the lower Option button (ANALYZE) to initiate rhythm analysis by the
FR2. If a shock is advised, the FR2 automatically charges, and prompts you to
press the Shock button.
After shock delivery, the H
EARTSTREAM FR2 returns to the advanced mode
display and monitors the patient’s heart rhythm. If a potentially shockable
rhythm is detected, the screen advises you to PRESS ANALYZE.
NOTE: If you do not press the lower Option button (labeled ANALYZE) to initiate
rhythm analysis when prompted, the H
EARTSTREAM FR2 does not analyze and
prompt if a shock is advised. It is important that you understand that entering the
advanced mode entails taking responsibility for these functions.
If the rhythm analysis results in a Shock Advised decision, the FR2 begins
charging, prompts you to press the Shock button, and displays a MANUAL
DISARM option at the top of the screen. If for any reason you want to cancel
the shock, press the upper Option button to disarm the FR2.
To return to non-manual, AED mode operation, turn the FR2 off then back on,
by pressing the On/Off button twice.
Using the Manual Charge Feature (M3860A only)
The manual charge feature is available only in the M3860A, and it
automatically includes display of the patient’s ECG. To enter the advanced
mode, press both Option buttons simultaneously. This brings up a screen that
includes a highlighted line at the top, labeled MANUAL with an arrowhead
pointing to the upper Option button, and another at the bottom, labeled
ANALYZE with an arrowhead pointing to the lower Option button.
Pressing the lower Option button (ANALYZE) provides user-initiated rhythm
analysis as described above. Pressing the upper Option button (MANUAL)
brings up a new screen. The highlighted top line is labeled MANUAL
CHARGE with an arrowhead pointing to the upper Option button.
If the ECG display shows that, in your expert clinical judgment, the patient
has a shockable rhythm, press the upper Option button (MANUAL
CHARGE). The H
EARTSTREAM FR2 will immediately charge for shock
delivery.
PRESS ANALYZE
0:43 2 HR XXX
ANALYZE >
MANUAL CHARGE>
MANUAL >
0:43 2 HR XXX
ANALYZE >

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M3860A and is the answer not in the manual?

Agilent Technologies M3860A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM3860A
CategoryMedical Equipment
LanguageEnglish

Related product manuals