Using Advanced Mode Features
M3860A and M3861A H
EARTSTREAM FR2 Defibrillator 6-13
6
As soon as charging begins, the screen message changes to CHARGING,
STAND CLEAR, and the label for the arrowhead pointing to the upper Option
button changes to MANUAL DISARM.
The FR2 beeps while it is charging. When the beeping changes to a
continuous tone and the Shock button light flashes, press the Shock button to
deliver a shock. However, if the ECG display shows that the patient’s rhythm
has changed to a non-shockable rhythm, press the upper Option button to
disarm the H
EARTSTREAM FR2.
After shock delivery, the H
EARTSTREAM FR2 returns to the initial advanced
mode screen. To return to non-manual AED mode operation, turn the FR2 off
then back on, by pressing the On/Off button twice.
MANUAL DISARM >
CHARGING
STAND CLEAR