6-10 Setup and Advanced Mode Features
the sending FR2 displays READY TO SEND and prompts you to check
the receiving FR2.
8. Clock settings are automatically transferred as soon as the infrared ports
are correctly aligned.
Using Advanced Mode Features
The HEARTSTREAM FR2 provides an advanced mode that allows responders
who are appropriately trained to override the programmed FR2 protocol and
take responsibility for certain aspects of the operating sequence used by the
FR2 to treat the patient.
As described earlier in this chapter, the factory default setup of the FR2 must
be modified to provide access to advanced mode features. This requires use of
the administration function of the M3864A Training & Administration Pack.
If you are an expert user authorized by your Medical Director to modify setup,
hold down both the Option buttons while installing the Training &
Administration Pack in the FR2, then select SETUP. Then select MODIFY
SETUP from the SETUP menu. Select ADVANCED from the third menu of
the MODIFY SETUP menu.
Using the upper Option button, scroll through the available settings for
ADVANCED. The advanced mode options available are based on the FR2
model used. For the M3860A, the user can select ANALYZE or CHARGE.
For the M3861A the user can select only ANALYZE. (Detailed directions for
use are supplied with the Pack.)
MODIFY SETUP
RETURN
ADVANCED OFF
CPR PROMPT LONG
PROMPT INTERVALS
MONITOR 1.0
ADVANCED USE 0.5
ADMINISTRATION
SETUP
CLOCK
REMOVE BATTERY
TO QUIT
SETUP
RETURN
RECEIVE SETUP
READ SETUP
MODIFY SETUP
SEND SETUP
WRITE SETUP

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M3860A and is the answer not in the manual?

Agilent Technologies M3860A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM3860A
CategoryMedical Equipment
LanguageEnglish

Related product manuals