Using Setup Features
6-8 Setup and Advanced Mode Features
Receiving Setup
This method uses the infrared communication feature of the HEARTSTREAM
FR2 to receive setup directly from one H
EARTSTREAM FR2 to another (which
must have the Training & Administration Pack installed in it) or from a
computer running C
ODERUNNER software. (See instructions provided with
C
ODERUNNER.) To receive setup from another FR2, follow these steps:
1. Locate the infrared communication port on each H
EARTSTREAM FR2 and
line them up with one another, so that the infrared “eye” in each one has
an uninterrupted view of the “eye” in the other. (See the diagram on the
inside front cover.) The two devices should be no more than 1 meter apart.
2. Make sure the “sending” H
EARTSTREAM FR2 has the Training &
Administration Pack installed and is ready to send. (See the M3864A
H
EARTSTREAM FR2 Training & Administration Pack Reference Guide for
instructions.)
3. Select RECEIVE SETUP from the setup menu:
4. A new screen comes up. Until the two H
EARTSTREAM FR2 devices are
properly positioned, the screen displays READY TO RECEIVE and
prompts you to check the sending FR2.
5. Setup data are automatically transferred as soon as the infrared ports are
correctly aligned.
6. If you select EXIT before the transfer is complete, the revised setup will
not be received. When the transfer is complete, the screen on the
“receiving” FR2 displays a SETUP COMPLETE message. Your
H
EARTSTREAM FR2 immediately uses the new setup.
Receiving setup from a computer running C
ODERUNNER software is
discussed in the directions for use of C
ODERUNNER.
Reading Setup
This method copies setup data from a data card to your HEARTSTREAM FR2.
To read the setup, follow these steps:
1. Insert the data card in the data card tray and install the loaded tray into the
data card slot in the H
EARTSTREAM FR2.
2. Select READ SETUP from the setup menu.
SETUP
RETURN
RECEIVE SETUP
READ SETUP
REVIEW SETUP
RECEIVE SETUP
EXIT
READY TO RECEIVE
CHECK SENDER
SETUP
RETURN
RECEIVE SETUP
READ SETUP
REVIEW SETUP

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M3860A and is the answer not in the manual?

Agilent Technologies M3860A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM3860A
CategoryMedical Equipment
LanguageEnglish

Related product manuals