Running the Battery Insertion Selftest
2-4 Preparing Your H
EARTSTREAM FR2 for Use
3. Use the lower Option button to select the next part you want to change,
and repeat the process, until all parts of the time and date have been set.
4. When you have made all the changes, move the highlight bar to RETURN
and press the upper Option button to go back to the main menu screen.
NOTE: New clock settings are used by the FR2 as soon as you set them. The clock
time display is updated each minute this screen is displayed. The clock seconds,
although not displayed, are set to 00 when you move the highlight bar out of the time
settings.
NOTE: If the battery is removed from the FR2 for more than two hours, the clock
settings will be lost and must be reset.
Running the Battery Insertion Selftest
Except in an emergency, it is recommended that you run this selftest every
time you change the battery.
The selftest has two parts. The first part tests the FR2 circuitry and memory.
The second part is interactive and requires you to respond to prompts in order
to make sure the display, buttons, lights, speaker, and beeper of the
H
EARTSTREAM FR2 are working properly. (See Chapter 4, Troubleshooting,
Testing, and Maintaining Your H
EARTSTREAM FR2, for details about this
selftest.)
To run the selftest:
1. Make sure the FR2 has not been in use for at least 5 minutes, and that the
defibrillation pads are not connected to the device.
2. Insert the battery into the battery port. The first screen displayed has RUN
SELFTEST highlighted.
3. Press the upper Option button to activate the test.
4. OR, make no selection for 10 seconds, and the selftest will start
automatically.
When the first part of the selftest is successfully completed, the screen
displays a message that the test has passed, and then automatically starts the
RUN SELFTEST
REVIEW INCIDENT
NO DATA CARD
GOOD BATTERY
NEXT
IN EMERGENCY
PRESS OFF TO QUIT

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M3860A and is the answer not in the manual?

Agilent Technologies M3860A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM3860A
CategoryMedical Equipment
LanguageEnglish

Related product manuals