Placing and Securing the HEARTSTREAM FR2
M3860A and M3861A H
EARTSTREAM FR2 Defibrillator 2-5
2
interactive part of the selftest. (See Chapter 4, Troubleshooting, Testing, and
Maintaining Your H
EARTSTREAM FR2, for details about the interactive part of
the test.)
When the selftest is complete, the H
EARTSTREAM FR2 automatically turns off
and returns to standby mode. The Status Indicator shows a flashing black
hourglass. In the standby mode, the FR2 is ready for use at any time, simply
by pressing the On/Off button to turn it on.
Placing and Securing the HEARTSTREAM FR2
Place the HEARTSTREAM FR2 in an accessible area with the Status Indicator
easily visible. Useful accessories for placing and securing the FR2 include a
carrying case, which is suitable for use with a wall mount bracket. (See
Appendix A for a list of accessories.)
With the battery installed and the FR2 stored in appropriate environmental
conditions, the H
EARTSTREAM FR2 performs detailed periodic selftests, to
make sure that it remains ready for use. (See Appendix B for the
environmental storage specifications.)
While the FR2 is in the standby mode, the Status Indicator shows the flashing
black hourglass unless the periodic selftests detect a problem. If a problem is
detected, the Status Indicator will show a flashing red
X or a solid red X and
the FR2 will “chirp” to alert you to the need for troubleshooting. (See Chapter
4, Troubleshooting, Testing, and Maintaining Your H
EARTSTREAM FR2, for
instructions.)
SELFTEST PASSED
REV: 01EM30 0201
NO DATA CARD
SN 000000001
IN EMERGENCY
PRESS OFF TO QUIT