Setup Overview
M3860A and M3861A H
EARTSTREAM FR2 Defibrillator 6-5
6
PAU SE K EY OFF,
MONITOR,
ALWAYS
OFF Enables or disables user-initiated CPR pause in the
automatic protocol. The length of the pause is defined by the
CPR Timer setting. When an Advanced mode feature
(ANALYZE or CHARGE) is enabled and accessed, the
Pause key is disabled.
OFF — disables availability of user-initiated pause.
MONITOR — enables user-initiated pause only during
FR2 monitoring of patient rhythm.
ALWAYS — enables user-initiated pause any time except
when the device is already paused.
If enabled, the Pause Key is
accessed by pressing the
lower Option key indicated
by an arrow on the FR2
display, as shown in the
sample screen:
RESUME KEY ON, OFF OFF Enables (ON) or disables (OFF) user-initiated interruption
of the CPR pause and return to analyzing. If either the CPR
Timer or the NSA Action setting is programmed to 1.5
minutes or longer, the Resume Key setting is automatically
enabled (ON).
If enabled, the Resume Key is
accessed by pressing the
lower Option key indicated
by an arrow on the FR2
display, as shown in the
sample screen:
ADVANCED USE
PROMPT
INTERVALS
(minutes)
0.5, 1.0, 1.5,
2.0, 2.5, 3.0
0.5 Sets the interval for “Press to Analyze” prompts provided
during advanced mode operation.
PARAM ET ER SETTINGS DEFAULT DESCRIPTION
MONITORING
ATTEND TO PATIENT
0:43 2 HR XXX
PAUSE FOR CPR >
PAUSED
ATTEND TO PATIENT
0:43 2
RESUME ANALYZING >