Setup Overview
M3860A and M3861A H
EARTSTREAM FR2 Defibrillator 6-3
6
PROTOCOL
TIMEOUT
(minutes)
0.5, 1.0, 1.5,
2.0, 2.5, 3.0,
3.5, =(infinite)
1.0 Sets the time interval used to determine if a delivered shock
should be counted as part of the current Shock Series.
CPR TIMER
(minutes)
0.5, 1.0, 1.5,
2.0, 2.5, 3.0
1.0 Sets the length of the CPR pause period that automatically
starts when:
• a Shock Series is completed; or
• the Pause Key (if enabled) is pressed; or
• an NSA decision is made, the NSA CPR pause is
enabled, and the conditions for using the CPR Timer
setting for the NSA CPR pause period are met (see
NSA Action); or
• the Shock button is not pressed for 30 seconds after the
FR2 is armed in the AED mode, or
• continuous artifact is encountered during rhythm
analysis.
After the CPR pause, the FR2 returns to automatic rhythm
analysis.
NSA ACTION
(minutes)
MONITOR,
0.5, 1.0, 1.5,
2.0, 2.5, 3.0
MONITOR Sets how the FR2 behaves following a NO SHOCK
ADVISED (NSA) decision:
MONITOR — directs the FR2 to monitor the patient’s
ECG following an NSA decision and to prompt the user
periodically to provide CPR. The interval for CPR
prompting is set by the Monitor Prompt Interval.
TIME SETTING — directs the FR2 to provide a CPR
pause period following an NSA decision (NSA CPR
Pause). If shocks have been delivered in the current
Shock Series (e.g., the NSA decision follows a shock),
the length of the CPR pause is defined by the CPR Timer
setting. Otherwise (e.g., the patient’s initial monitored
rhythm is non-shockable), the length of the CPR pause is
defined by the NSA Action time setting.
PAR AM ETE R SETTINGS DEFAULT DESCRIPTION