Troubleshooting Guide
4-4 Troubleshooting, Testing, and Maintaining Your H
EARTSTREAM FR2
Flashing
red X:
Audio signal:
chirping
• Battery power is low.
•The H
EARTSTREAM FR2 has been
stored outside the recommended
temperature range.
• An error has been detected as part
of the selftest.
• The FR2 has been unable to
perform its daily selftests.
• Remove and reinstall the battery and
run a battery insertion selftest. A
screen prompt will tell you if
H
EARTSTREAM FR2 has been stored
outside the recommended temperature
range. See Appendix B for
recommended range.
• Remove and reinstall the battery and
perform the battery insertion selftest. If
it fails, install a new battery and repeat
the test. If it fails again, do not use the
H
EARTSTREAM FR2.
• Make sure defibrillation pads are not
attached to the FR2.
Flashing or
solid red X
Audio signal:
chirping
Screen prompt
(displayed for
10 seconds
before FR2
turns off):
NOT READY
FOR USE
or
SELFTEST
FAILED
The HEARTSTREAM FR2 performs
selftests every time you turn it on or
insert a battery, and periodically
while it is in standby mode. A test
revealed a failure or error.
• Unplug the pads connector from the
H
EARTSTREAM FR2, if connected.
• Remove and reinstall the battery and
check the results of the battery
insertion selftest. If it fails, install a
new battery and repeat the test. If it
fails again, do not use the
H
EARTSTREAM FR2.
NOTE: You can stop the tests and use the
H
EARTSTREAM FR2 as soon as you see the
Status Indicator change to the flashing
black hourglass. Simply press the On/Off
button to stop the test and put the FR2
into standby mode. The FR2 is then ready
for use.
STATUS
INDICATOR
OTHER
SIGNAL
POSSIBLE CAUSE RECOMMENDED ACTIONS