Maintenance
M3860A and M3861A H
EARTSTREAM FR2 Defibrillator 4-17
4
Cleaning the HEARTSTREAM FR2
The outside of the HEARTSTREAM FR2, including the defibrillator pads
connector port, can be cleaned with a soft cloth dampened in one of several
appropriate cleaning agents (see list below). The following guidelines include
some important reminders:
Do not immerse the HEARTSTREAM FR2 in fluids.
Make sure a battery (or the Training & Administration Pack) and a data
card tray are installed when cleaning the H
EARTSTREAM FR2, to keep
fluids out of the device.
Do not use abrasive materials, cleaners, strong solvents such as acetone or
acetone-based cleaners, or enzymatic cleaners.
Clean the HEARTSTREAM FR2 and the connector socket with a soft cloth
dampened with one of the cleaning agents listed below.
Isopropyl alcohol (70% Solution)
Soapy water
Chlorine bleach (30 ml/l water)
Ammonia based cleaners
Glutaraldehyde-based cleaners
Hydrogen peroxide
CAUTION: Do not immerse any portion of the HEARTSTREAM FR2 in water or other fluids. Do not allow
fluids to enter the H
EARTSTREAM FR2. Avoid spilling any fluids on the HEARTSTREAM FR2 or
accessories. Spilling fluids into the H
EARTSTREAM FR2 may damage it or present a fire or shock
hazard. Do not sterilize the HEARTSTREAM FR2 or accessories.