Safety Considerations
M3860A and M3861A H
EARTSTREAM FR2 Defibrillator 5-3
5
CAUTION Hazardous electrical output. The HEARTSTREAM FR2 is for use only by qualified personnel.
CAUTION Do not immerse any portion of the HEARTSTREAM FR2 in water or other fluids. Do not allow fluids
to enter the H
EARTSTREAM FR2. Avoid spilling any fluids on the HEARTSTREAM FR2 or accessories.
Spilling fluids into the H
EARTSTREAM FR2 may damage it or present a fire or shock hazard. Do not
sterilize the H
EARTSTREAM FR2 or accessories.
SAFETY LEVEL POSSIBLE IMPROPER DEVICE PERFORMANCE
WARNING The HEARTSTREAM FR2 runs a selftest whenever a new battery is installed and
automatically runs periodic selftests. These tests are designed to check that the
H
EARTSTREAM FR2 is ready to use. However, the testing cannot assure proper performance
if the H
EARTSTREAM FR2 was abused or damaged after the last selftest.
WARNING Prolonged or aggressive CPR to a patient with defibrillation pads attached can damage
the pads. Replace the defibrillation pads if they are damaged during use or handling.
WARNING Using damaged or expired equipment or accessories may cause the H
EARTSTREAM FR2 to
perform improperly, and/or injure the patient or the user.
WARNING CPR rates significantly above the American Heart Association guidelines of 100
compressions per minute can cause incorrect or delayed analysis by the H
EARTSTREAM
FR2.
CAUTION The H
EARTSTREAM FR2 is designed to be used only with Agilent Technologies-approved
accessories. The H
EARTSTREAM FR2 will perform improperly if non-approved accessories are used.
CAUTION Follow all instructions supplied with the H
EARTSTREAM defibrillation pads. Use the defibrillation
pads before the expiration date shown on the package. Do not reuse the defibrillation pads.
Discard them after use.
CAUTION: Aggressive handling of the pads in storage or prior to use can damage the pads. Discard the
defibrillation pads if they become damaged
CAUTION Follow all instructions supplied with the H
EARTSTREAM M3863A battery pack. Install the battery
before the expiration date shown on the battery.
CAUTION The H
EARTSTREAM FR2 was designed to be sturdy and reliable for many different field use
conditions. However, excessively rough handling can result in damage to the H
EARTSTREAM FR2 or
its accessories. Inspect the unit and accessories periodically according to instructions.
SAFETY LEVEL POSSIBLE SHOCK OR FIRE HAZARD, OR EXPLOSION

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M3860A and is the answer not in the manual?

Agilent Technologies M3860A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM3860A
CategoryMedical Equipment
LanguageEnglish

Related product manuals