Recording Incident Data
7-2 Data Management and Review
CAUTION: The HEARTSTREAM FR2 is designed to be used only with Agilent Technologies-approved
accessories. The HEARTSTREAM FR2 may perform improperly if non-approved accessories are used.
To insta ll a data ca r d:
1. Make sure the data card is clean and dry.
2. Load the data card into its plastic tray, with the tray’s “tongue” fitting
over the matching yellow area on the data card. The label on the card
should face up. The label has an arrow indicating which side to insert
into the data card port.
3. Make sure the FR2 is off (in standby mode), or that the battery has
been removed.
4. Hold the loaded tray by its handle and gently insert the tray into the
data card port on the right side of the H
EARTSTREAM FR2. Push the
tray all the way into the port. Do not force the tray into the port. If the
tray is hard to insert, remove it and make sure that the arrow label is
face up and pointing toward the data card port.
The data card will automatically record incident data the next time the
H
EARTSTREAM FR2 is turned on.
To avoid running out of data card space during an incident, it is
recommended that each data card be used to record the information for
only one incident and be replaced after each use of the H
EARTSTREAM
FR2.
If you record information from more than one incident on a data card, it is
important to review how much time is left on the used data card before
recording a new incident. To do this, load the data card into the data card
tray, insert the tray in the FR2, then remove and reinstall the battery. The
first screen displayed shows how much recording time remains on the
card.
NOTE: During an incident, if for any reason you replace the battery or turn off the
H
EARTSTREAM FR2 for less than five minutes, the FR2 considers this to be a
“continued use” situation, and:
the information stored about the incident is saved,

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M3860A and is the answer not in the manual?

Agilent Technologies M3860A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM3860A
CategoryMedical Equipment
LanguageEnglish

Related product manuals