If condensation is found in tubes during a run
Agilent SureCycler 8800 Setup and User’s Guide 59
If condensation is found in tubes during a run
Condensation will appear naturally at the end of a run and has no
deleterious effects.
Spin the liquid back to the bottom of the tube and continue with the
post PCR
st
eps.
The hot top is disabled.
Enable the hot top using the instructions in “Hot Top Setting” on
page 32.
If the sealed microplate loses sample volume
Check that the microplate/sealing system is of good quality.
Ensure a compression mat is used during the PCR run to keep the
sea
ling film adhered to the plate.
If USB devices including memory stick and keyboard not
recognized
Reboot the cycler.
If cycler does not turn on
Check that the power is on at supply and the power cord is connected.
Check fuses at rear of instrument above power inlet.