Maintenance      5
Agilent U1251A/U1252A User’s and Service Guide 107
Charging the Battery
This meter is powered by a 7.2 V NiMH rechargeable battery. 
Use the specified accessory, the 24 V DC adaptor to charge 
the battery. Remember never to turn the rotary switch of 
the meter when its battery is being charged. Use the follow-
ing procedures to charge the battery:
1 Remove and disconnect the test leads from the meter.
2 Turn the rotary switch to the   position. Connect 
the power cord to the DC adapter.
3 Plug the Red (+)/ Black (- ) banana terminals of the DC 
adapter to the   and the “COM” terminals 
respectively.  The DC adaptor can be replaced with a DC 
power supply in order to set the 24 V DC output and the 
over current limitation to a value more than 0.5 A. 
Ensure that the polarity of the connection is correct.
4 The primary display indicates “bAt” and the ‘SbY” flashes 
on the secondary display and a short tone sounds to 
remind you whether to charge the battery or not. Press 
SHIFT to start charging the battery, or the meter will 
automatically start the self- test after the 24 V supply is 
applied. We recommend not to charge the battery if its 
capacity is over 90%.
Never discharge the battery by shorting it or by reverse polarity in any 
subjects. Make sure the battery is the rechargeable type before charging 
it. Never rotate the rotary switch when the battery is being charged as DC 
at 24 V is being applied to the charging terminals.
For the battery charger, the mains supply voltage should not fluctuate by
plus or minus 10%. 
Condition Battery Voltage Proportional Percentage
Trickle (SBY) 6.0 V ~ 8.2 V 0% ~ 100%
Under charging 7.2 V ~ 10.0 V 0% ~ 100%