EasyManua.ls Logo

Agilent Technologies U1251A - Calibration Process

Agilent Technologies U1251A
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
142 Agilent U1251A/U1252A User’s and Service Guide
7 Performance Tests and Calibration
Now you can use 1234 as the security code. If you want to
enter a new security code, see “To Change the Instrument
Calibration Security Code from the Front Panel” on
page 140. Make sure you record the new security code.
Calibration Process
To complete a full instrument calibration use following
general procedures:
1 Read “Test Considerations” on page 128.
2 Perform the verification tests to characterize the
instrument (incoming data).
3 Unsecure the instrument for calibration (see “Calibration
Security” on page 138).
4 Perform the adjustment procedures (see “Adjustments
Consideration” on page 144).
5 Secure the instrument against calibration.
6 Note the new security code and calibration count in the
instrument's maintenance records.
NOTE
Make sure to quit the Adjustment Mode before turning off the instrument.

Table of Contents

Other manuals for Agilent Technologies U1251A

Related product manuals