1 ms Peak Hold
This function allows the measurement of half- cycle peak 
voltage for the the analysis of components such as power 
distribution transformers and power factor correction capac-
itors. The peak voltage obtained can be used to determine 
the crest factor:
Crest factor = Peak value/True RMS value
1 Press   for more than 1 second to toggle 1 ms Peak 
Hold mode ON / OFF.
2 Press   to scroll through maximum and minimum 
peak readings.   MAX indicates the maximum peak, 
while   MIN indicates the minimum peak.
3 Press   or   for more than 1 second to exit this 
mode. 
4 According to the measurements in Figure 22 on the next 
page, the crest Factor will be 2.5048/1.768 =1.416.
• If the reading is "OL", press  to change the measuring range and 
to re-start the peak-recording measurement.
• If you need to re-start the peak recording, press