130 Agilent U1251A/U1252A User’s and Service Guide
7 Performance Tests and Calibration
Use the Performance Verification Tests to verify the mea-
surement performance of the instrument. The performance
verification tests use the instrument’s specifications listed in
the U1251A/U1252A Data Sheet.
The performance verification tests are recommended as
acceptance tests when you first receive the instrument. The
acceptance test results should be compared against the one
year test limits. After acceptance, you should repeat the per-
formance verification tests at every calibration interval.
If the multimeter fails the performance verification tests,
adjustment or repair is required. Adjustment is recom-
mended at every calibration interval. If adjustment is not
made, you must establish a ‘guard band’, using not more
than 80% of the specifications, as the verification limits.
NOTE
Make sure you have read the “Test Considerations” on page 128 before
doing the performance verification tests.
Performance Verification Tests

Table of Contents

Other manuals for Agilent Technologies U1251A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1251A and is the answer not in the manual?

Agilent Technologies U1251A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1251A
CategoryMultimeter
LanguageEnglish

Related product manuals