Performance Tests and Calibration 7
Agilent U1251A/U1252A User’s and Service Guide 147
8 Press to start the adjustment. The CAL flashes in
the secondary display to indicate that the calibration is in
progress.
For an adjustment that is successful, the secondary display
will briefly show PASS. An adjustment failure is indicated by
a long beep and a calibration error number appears in the
secondary display. The primary display remains at the
current Cal Item. Check the input value, range, function, and
entered adjustment value to fix the problem. Repeat the
adjustment procedure.
9 Repeat steps 1 through 8 for each adjustment point.
10 Verify the adjustments using the “Performance Verification
Tests on page 130. Verify the adjustments using Table 6: