Performance Tests and Calibration 7
Agilent U1251A/U1252A User’s and Service Guide 149
Step Test Function Cal Range Input Cal Item
U1251A U1252A
3 Press button to go to the
V mode
5 V 0.3 V, 1 kHz N/A 0.3000 V
3 V, 1 kHz N/A 3.0000 V
3 V, 50 kHz N/A 3.0000 V
50 V 3 V, 1 kHz N/A 03.000 V
30 V, 1 kHz N/A 30.000 V
30 V, 50 kHz N/A 30.000 V
500 V 30 V, 1 kHz N/A 030.00 V
300 V, 1 kHz N/A 300.00 V
300 V, 50 kHz N/A 300.00 V
1000 V 30 V, 1 kHz N/A 0030.0 V
300 V, 1 kHz N/A 0300.0 V
300 V, 50 kHz N/A 0300.0 V
4 Turn the rotary switch to the
mV position
Short Dual Banana Plug
with copper wire
short between 2
terminals
SHort SHort
50 mV 30 mV 30.000 mV 30.000 mV
500 mV 300 mV 300.00 mV 300.00 mV
1000 mV 1000 mV 1000.0 mV 1000.0 mV