6.3.2  Measurement Calibration - Low Scale 
Note:  Measurement calibration can be done at any output frequency setting as 
only one set of calibration coefficients is used for measurement 
calibration.  However, for best results, calibrate the measurements at the 
frequency, which is used in most applications, e.g. 50 Hz, 60 Hz (except 
model 2003RP-AV) or 400 Hz. 
Refer to Table 6-1 for correct load resistors based on model number. 
Low Scale Voltage Measurement Calibration 
1.  Connect the test equipment to the power source as shown in Figure 6-2.  Do not 
connect any load during this step.  Voltage calibration must be done under no load 
conditions. 
2.  Select the Low Voltage Range and program the output voltage to 10 VAC. 
3.  Put the 2003RP in CAL mode by pressing the recessed CAL button. 
4.  Use the SELECT key to select the Voltage Measurement calibration mode.  This mode is 
indicated by the FREQ LED. 
5.  Use the left control knob to adjust the voltage reading shown on the right hand 7 
segment LED until it reads as close as possible to the reference reading as displayed on 
the external DMM2. 
 
Figure 6-2: Test Equipment Hook-up for Current and Power Measurement Calibration 
Low Scale Current Measurement Calibration 
1.  Connect the test equipment to the power source as shown in Figure 6-2. 
2.  Apply the resistive load (see Table 6-1 for model and load) to the output terminals.  Make 
sure external DMM1 used for current measurement is across the shunt. 
3.  The output voltage must be programmed to 10.0 volts. 
4.  Use the SELECT key to select the Current Measurement calibration mode. This mode is 
indicated by the I RMS LED. 
5.  Use the left control knob to adjust the current reading shown on the right hand 7 segment 
LED until it reads as close as possible to the reference reading indicated by the DMM 
2003RP
AC SOURCE
 Pow er A  1
 Sense A  2
 Pow er B  3
 Sense B  4
 Pow er C  5
 Sense C  6
 Pow er N  7
 Sense N  8
DMM 2
AC VOLT
DMM 1
AC VOLT
0.1 ohm Shunt
Load Resistor