2 Specifications
All specifications at 25 5 C and 400 Hz output unless noted otherwise.
2.1 Electrical
Input
Parameter
Specification
Line Voltage:
107/120 V
AC
± 10%
208/230 V
AC
± 10%
Line Current:
35 A
RMS
maximum, line voltage = 96 V
AC
, full-load,
PF = 0.6, Efficiency = 75%, (typical)
Inrush Current:
185 A
PK
at 253 V
AC
input line
Input power factor:
0.6 at full-load and 120 V
AC
input
Line Frequency
1
:
50/60 Hz ± 5%
Efficiency:
75% typical (depending on line and load)
Hold-Up Time:
10 ms (with no effect on output)
Isolation Voltage:
1350 V
AC
, input to chassis
2300 V
AC
, input to output
Output
Parameter
Specification
Outputs
Phases:
3
Mode:
AC
Coupling:
AC
Voltage
2003RP:
0 to 135 V or 0 to 270 V
(see Section 3.7)
2003RP-AV:
0 to 115 V or 0 to 230 V
or
0 to 115 V only (factory set)
2003RP-HV:
0 to 156 V or 0 to 312 V
(see Section 3.7)
Resolution:
0.1 volt
Accuracy
1
:
± 0.1% of full scale, 16 Hz - 100 Hz
± 0.2% of full scale, 100 Hz - 2000 Hz
Applies from 10 to 100% of voltage range.
1
Supplementary specification applies outside indicated frequency ranges.