In the event the power source does not pass the functional test, refer to the calibration 
procedure in Section 6 or call California Instrument‟s customer satisfaction department for 
further assistance. 
 
 
Figure 3-7: Test Setup 2003RP. 
 
3.9  Other Modes of Operation 
When the 2003RP is shipped from the factory, it is configured for the Constant Current Mode 
of operation.  This mode means that if the load current exceeds the programmed Current 
Limit value, the output voltage will drop and the current will continue at the programmed 
value. 
The 2003RP can be configured to operate in the Constant Voltage mode.  In this mode of 
operation, if the load current exceeds the Current Limit value, the output voltage will default 
with the output relay open.  The Constant Voltage mode will only work if the default voltage is 
set for 5.0 volts.  If the default is any other voltage, the power source may indicate a Fault at 
power up. 
To configure the 2003RP, remove the input line voltage connection from the rear panel.  
Remove the top cover.  Install jumpers W1 and W2 on controller assembly, A1, and jumpers 
W1, W3, W4 and W6 on the Phase B/C assembly, A2.  Both assemblies are on the front 
panel.  Reinstall the top cover. 
L N G
GROUND
NEUTRAL
LINE
1 2 3 4 5 6 78
Load
To Voltmeter