4.1.2  Control knobs 
 
 
        Counter      Clockwise 
       clockwise   
 
DECREASE      INCREASE   
 
Figure 4-3:  Control Knob 
There are two control knobs located below the LED displays which are used to change setup 
parameters for voltage, frequency and current limit. The SELECT button selects the function 
of the right knob. The right knob will control either the frequency or the current limit as 
indicated by the indicator above the right LED. The left knob always controls the voltage. 
Setting changes always affect all three output phases simultaneously. 
4.1.3  Buttons 
There are four function buttons for the Output Voltage Range, Output State, Select Mode, 
and Phase Select.  The following is a description of these buttons: 
KEY  DESCRIPTION   
RANGE  The RANGE button is used to change the voltage range 
between the low range (0 to 135 or 115 volts) and high 
range (0 to 270 or 230 volts).  The LED above the switch will 
light to indicate the high voltage range selection.  The output 
voltage will be reset to zero voltage after a range change.  If 
the 2003RP-AV is configured for a single voltage range, 
this button is disabled. 
The same button may be used to put the 2003RP in the 
AUTO voltage range change mode. This button toggles 
from Lo, Auto, Hi and back to Lo. The AUTO LED will 
illuminate when the AUTO range mode is enabled.  While in 
AUTO mode, the 2003RP will switch to the high range 
whenever a value larger than the low range limit is 
programmed. Whenever a value below the low range limit is 
programmed, the unit will switch to the low voltage range.  
The output is momentarily interrupted when a range change 
occurs. 
To take the unit out of AUTO mode, press the RANGE 
button again. 
OUTPUT  The OUTPUT button will toggle the output relays to connect 
or disconnect the outputs.  The LED above the button will 
light when the output relays are closed. No output voltage 
will be present when the OUTPUT button is off despite the 
level of voltage programmed.