7.3.8 Can’t Program AC Power System on GPIB or RS232
If the power source does not respond to GPIB or RS232 programming, the following items
may be at fault:
CAUSE
SOLUTION
The power source unit address is
incorrect.
Set correct address. See section 8.3.
GPIB or RS232 cable is loose at power
source rear panel.
Check connection, tighten jack screws.
The GPIB/RS232 selector switch is in the
wrong position.
Set DIP switch to desired function. Refer
to Section 5.9
RS232/GPIB Assembly has failed.
Replace the RS232/GPIB Assembly.
7.4 Module Removal
Figure 7-1 shows the location of the internal modules and assemblies.