Chapter 6, COMPUTER OPERATION
6-14
synchronous time constant control is turned off, the fast analog output mode is
turned on (FASTMODE 1), and the output filter slope is set to 12 dB/octave if it had
previously been 18 or 24 dB/octave.
NNBUF [n] Noise Buffer Length control
The value of n sets the noise buffer length according to the following table:
n Function
0 Noise buffer off
1 1 second noise buffer
2 2 second noise buffer
3 3 second noise buffer
4 4 second noise buffer
TC [n]
TC. Filter time constant control
The value of n sets the output filter time constant in accordance with the following
table:
N/A means that n is an illegal value under these conditions.
The TC. command is only used for reading the time constant, and reports the current
setting in seconds. Hence if a TC 12 command were sent and the noise measurement
mode was turned off, TC would report 12 and TC. would report 1.0E-01, i.e. 0.1 s or
100 ms.
SYNC [n] Synchronous time constant control
n Effect
0 Synchronous time constant disabled
1 Synchronous time constant enabled