EasyManuals Logo

Ametek 7230 User Manual

Ametek 7230
171 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #105 background imageLoading...
Page #105 background image
Chapter 6, COMPUTER OPERATION
6-19
6.7.06 Internal Oscillator
OA[.] [n] Oscillator amplitude control
In fixed point mode n sets the oscillator amplitude in microvolts rms. The range of n
is 0 to 5000000 representing 0 V to 5 V rms.
In floating point mode n sets the amplitude in volts.
ASTART[.] [n] Oscillator amplitude sweep start amplitude
Sets the start amplitude for a subsequent sweep of the internal oscillator amplitude,
in the range 0 to 5.000 V
In fixed point mode, n is in microvolts rms, and in floating point mode n is in volts
rms
ASTOP[.] [n] Oscillator amplitude sweep stop amplitude
Sets the stop amplitude for a subsequent sweep of the internal oscillator amplitude,
in the range 0 to 5.000 V
In fixed point mode, n is in microvolts rms and in floating point mode n is in volts
rms
ASTEP[.] [n] Oscillator amplitude sweep step size
Sets the amplitude step size for a subsequent sweep of the internal oscillator
amplitude, in the range 0 to 5.000 V
In fixed point mode, n is in microvolts rms and in floating point mode n is in volts
rms
OF[.] [n] Oscillator frequency control
In fixed point mode n sets the oscillator frequency in mHz. The range of n is 0 to
120000000 or 250000000 representing 0 to 120 kHz or 250 kHz.
In floating point mode n sets the oscillator frequency in Hz. The range of n is 0 to
1.2E5 or 2.5E5
FSTART[.] [n] Oscillator frequency sweep start frequency
Sets the start frequency for a subsequent sweep of the internal oscillator frequency,
in the range 0 to 120 or 250 kHz.
In fixed point mode, n is in millihertz, and in floating point mode n is in hertz.
FSTOP[.] [n] Oscillator frequency sweep stop frequency
Sets the stop frequency for a subsequent sweep of the internal oscillator frequency,
in the range 0 to 120 or 250 kHz.
In fixed point mode, n is in millihertz and in floating point mode n is in hertz.
FSTEP[.] [n
1
n
2
] Oscillator frequency sweep step size and type
The frequency may be swept either linearly or logarithmically, as specified by
parameter n
2
. The step size is specified by parameter n
1
.
Log sweep n
2
= 0
In fixed point mode, n
1
is the step size in thousandths of a percent.
In floating point mode n
1
is in percent. The range of n
1
is 0 to 100.000%

Table of Contents