Chapter 5, WEB CONTROL PANEL OPERATION 
5-28 
5.3.13 Oscillator: Sweep Control 
 
Figure 5-31, Oscillator Panel, Sweep Controls 
The internal oscillator output in the model 7230 can be swept in frequency or 
amplitude, or in both parameters, using these controls. 
Time/Step 
This control defines the time per point during a frequency or amplitude sweep 
Start 
This button is used to start or stop a frequency and/or amplitude sweep. The Armed 
checkbox in the required sweep type controls must also be checked in order to run a 
sweep. 
Frequency Sweep group 
  Start frequency 
This control defines the start frequency from which the instrument will start a 
frequency sweep. 
  Stop frequency 
This control defines the end frequency for the frequency sweep 
  Step size 
This control defines the amount by which the oscillator frequency is changed at 
each step. Depending on the sweep law selected (linear or logarithmic) it is set 
either in hertz, or as a percentage of the present frequency. If Start Frequency is 
greater than Stop Frequency then the output frequency will decrease with time. 
  Armed  
If this box is checked then the defined sweep will be run when the Start button is 
clicked. 
  Linear 
When this radio button is selected, all frequency steps are equal and the 
frequency increases linearly with time.