Chapter 3, TECHNICAL DESCRIPTION 
3-9 
modulation is applied to the angle of the grating (in optical spectroscopy) or to the 
applied magnetic field (in magnetic resonance spectroscopy). Double beam 
spectroscopy is a further common example. 
In this signed scalar measurement the phase-shifter must be set, after removal of any 
zero errors, to maximize the X channel or the Y channel output of the demodulator. 
This is the only method that will give correct operation as the output signal passes 
through zero, and is also the best method to be used in an unsigned scalar 
measurement where any significant amount of noise is present. 
3.3.10 Internal Oscillator - General 
The model 7230, in common with many other lock-in amplifiers, incorporates an 
internal oscillator, which may be used to drive the experiment. However, unlike 
older instruments, the oscillator in the model 7230 is digitally synthesized with the 
result that the output frequency is extremely accurate and stable. The oscillator 
operates over the same frequency range as the lock-in amplifier, that is 1 mHz to 
120.0 or 250.0 kHz. The oscillator signal is available at the OSC OUT connector. 
3.3.11 Internal Oscillator - Update Rate 
The direct digital synthesis (DDS) technique generates a waveform at the DAC 
output, which is not a pure sinusoid, but rather a stepped approximation to one. This 
is then filtered by the buffer stage, which follows the DAC, to reduce the harmonic 
distortion to an acceptable level. The update rate is 2.0 MHz. 
3.3.12 Internal Oscillator - Frequency & Amplitude Sweeps 
The internal oscillator output may be swept in both frequency and amplitude. In both 
cases the sweeps take the form of a series of steps between starting and finishing 
values. Frequency sweeps may use equal increment step sizes, giving a linear change 
of frequency with time as the sweep proceeds, or may use step sizes proportional to 
the present frequency, which produces a logarithmic sweep. The amplitude sweep 
function offers only linear sweeps. 
A special form of the frequency sweep function is used to acquire lock when the 
instrument is operating in the virtual reference mode. When this "seek" sweep is 
activated, the oscillator starts at a user-specified frequency, which should be just 
below that of the applied signal, and increments until the calculated magnitude 
output is greater than 50%. At this point the sweep then stops and the virtual 
reference mode achieves lock, by continuously adjusting the internal oscillator 
frequency to maintain the Y channel output at zero. 
It is important to note that this type of phase-locked loop, unlike a conventional 
edge-triggered type using a clean reference, does not automatically re-acquire lock 
after it has been lost. Lock can be lost as a result of a signal channel transient or a 
phase reversal of the signal, in which case it may be necessary to repeat the lock 
acquisition procedure. However, if the measurement system is set up with sufficient 
precautions, particularly ensuring that the full-scale sensitivity is maintained at a 
suitable setting in relation to the signal level, then the virtual reference mode is 
capable of making signal recovery measurements which are not possible with other 
lock-in amplifiers. 
When virtual reference mode is in use, the signal at the OSC OUT connector is a 
sinusoid which is phase-locked to the signal. This cannot, of course, be used as a 
source for the measurement.