Chapter 6, COMPUTER OPERATION 
6-15 
SLOPE [n]  Output low-pass filter slope (roll-off) control 
When the fast analog output mode and noise measurement mode are turned off 
(FASTMODE 0 and NOISEMODE 0), the selection is according to the following 
table: 
n  Slope 
0  6 dB/octave 
1  12 dB/octave 
2  18 dB/octave 
3  24 dB/octave 
When: 
a)  The fast output filter mode is turned on, or 
b)  Noise measurement mode is turned ON (NOISEMODE 1) 
the selection is according to the following table: 
n  Slope 
0  6 dB/octave 
1  12 dB/octave 
6.7.04 Signal Channel Output Amplifiers 
XOF [n
1
 [n
2
]]  X channel output offset control 
n
1
  Selection 
0  Disables offset 
1  Enables offset facility 
The range of n
2
 is ±30000 corresponding to ±300% full-scale. 
YOF [n
1
 [n
2
]]  Y channel output offset control 
n
1
  Selection 
0  Disables offset facility 
1  Enables offset facility 
The range of n
2
 is ±30000 corresponding to ±300% full-scale. 
AXO  Auto-Offset 
The X and Y channel output offsets are turned on and set to levels giving zero X and 
Y channel outputs. Any changes in the input signal then appear as changes about 
zero in the outputs. 
EX [n]  Output expansion control 
Expands X and/or Y channel outputs by a factor of 10. This changes the analog 
outputs full-scale to ±10% if X or Y selected. The value of n has the following 
significance: 
n  Expand mode 
0  Off 
1  Expand X 
2  Expand Y 
3  Expand X and Y