Appendix E, ALPHABETICAL LISTING OF COMMANDS 
E-20 
The lowest five bits in n
2
 control the other RS232 parameters according to the 
following table: 
bit number  bit negated    bit asserted 
0    data = 7 bits    data = 8 bits 
1    no parity bit    1 parity bit 
2    even parity    odd parity 
3    echo disabled    echo enabled 
4    prompt disabled   prompt enabled 
RT[.]  Ratio output 
In integer mode the RT command reports a number equivalent to 1000×X/ADC1 
where X is the value that would be returned by the X command and ADC1 is the 
value that would be returned by the ADC1 command. 
In floating point mode the RT. command reports a number equivalent to X/ADC1. 
SEN [n] 
SEN.  Full-scale sensitivity control 
The value of n sets the full-scale sensitivity according to the following table, 
depending on the setting of the IMODE control: 
n      full-scale sensitivity 
    IMODE=0  IMODE=1  IMODE=2 
3    10 nV  10 fA  n/a 
4    20 nV  20 fA  n/a 
5    50 nV  50 fA  n/a 
6    100 nV  100 fA  n/a 
7    200 nV  200 fA  2 fA 
8    500 nV  500 fA  5 fA 
9    1 µV  1 pA  10 fA 
10    2 µV  2 pA  20 fA 
11    5 µV  5 pA  50 fA 
12    10 µV  10 pA  100 fA 
13    20 µV  20 pA  200 fA 
14    50 µV  50 pA  500 fA 
15    100 µV  100 pA  1 pA 
16    200 µV  200 pA  2 pA 
17    500 µV  500 pA  5 pA 
18    1 mV  1 nA  10 pA 
19    2 mV  2 nA  20 pA 
20    5 mV  5 nA  50 pA 
21    10 mV  10 nA  100 pA 
22    20 mV  20 nA  200 pA 
23    50 mV  50 nA  500 pA 
24    100 mV  100 nA  1 nA 
25    200 mV  200 nA  2 nA 
26    500 mV  500 nA  5 nA 
27    1 V  1 µA  10 nA 
Floating point mode can only be used for reading the sensitivity, which is reported in 
volts or amps. For example, if IMODE = 0 and the sensitivity is 1 mV the command 
SEN would report 18 and the command SEN. would report +1.0E-03. If IMODE was