EasyManua.ls Logo

Ametek 7230 - Page 163

Ametek 7230
171 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix E, ALPHABETICAL LISTING OF COMMANDS
E-25
n Trigger
0 Trigger generated once per curve, at start of curve
1 Trigger generated once per point
TRIGOUTPOL [n] Curve buffer trigger output polarity
The value of n defines the polarity of the trigger output generated when a buffer
acquisition is run in accordance with the following table:
n Trigger Output
0 Rising-edge
1 Falling edge
USBTERM [n] Set/read USB terminator
The value of n sets the USB terminator according to the following table:
n Function
0 A null character (ASCII 0)
1 A null character (ASCII 0), followed by a byte representing the value of the
status Byte (table 6-1), followed by a byte representing the value of the Overload
Byte (table 6-1)
VER Report firmware version
Causes the lock-in amplifier to respond with the firmware version number, as also
displayed on the Configuration menu.
VMODE [n] Voltage input configuration
The value of n sets up the input configuration according to the following table:
n Input configuration
0 Both inputs grounded (test mode)
1 A input only
2 -B input only
3 A-B differential mode
Note that the IMODE command takes precedence over the VMODE command.
VRLOCK [n] Virtual reference mode lock
The Seek option of the frequency sweep mode must be used before issuing this
command, for which the value of n has the following significance:
n Mode
0 Disables virtual reference mode
1 Enters virtual reference mode by enabling tracking of the signal frequency
X[.] X channel output
In fixed point mode causes the lock-in amplifier to respond with the X demodulator
output in the range ±30000, full-scale being ±10000.
In floating point mode causes the lock-in amplifier to respond with the X
demodulator output in volts or amps.

Table of Contents