EasyManuals Logo

Ametek 7230 User Manual

Ametek 7230
171 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
Chapter 6, COMPUTER OPERATION
6-10
DCCOUPLE [n] Input coupling control
The value of n sets the input coupling mode according to the following table:
n Coupling mode
0 AC coupled
1 DC coupled
SEN [n]
SEN. Full-scale sensitivity control
The value of n sets the full-scale sensitivity according to the following table,
depending on the setting of the IMODE control:
n full-scale sensitivity
IMODE=0 IMODE=1 IMODE=2
3 10 nV 10 fA n/a
4 20 nV 20 fA n/a
5 50 nV 50 fA n/a
6 100 nV 100 fA n/a
7 200 nV 200 fA 2 fA
8 500 nV 500 fA 5 fA
9 1 µV 1 pA 10 fA
10 2 µV 2 pA 20 fA
11 5 µV 5 pA 50 fA
12 10 µV 10 pA 100 fA
13 20 µV 20 pA 200 fA
14 50 µV 50 pA 500 fA
15 100 µV 100 pA 1 pA
16 200 µV 200 pA 2 pA
17 500 µV 500 pA 5 pA
18 1 mV 1 nA 10 pA
19 2 mV 2 nA 20 pA
20 5 mV 5 nA 50 pA
21 10 mV 10 nA 100 pA
22 20 mV 20 nA 200 pA
23 50 mV 50 nA 500 pA
24 100 mV 100 nA 1 nA
25 200 mV 200 nA 2 nA
26 500 mV 500 nA 5 nA
27 1 V 1 µA 10 nA
Floating point mode can only be used for reading the sensitivity, which is reported in
volts or amps. For example, if IMODE = 0 and the sensitivity is 1 mV the command
SEN would report 18 and the command SEN. would report +1.0E-03. If IMODE was
changed to 1, SEN would still report 18 but SEN. would report +1.0E-09
AS Perform an Auto-Sensitivity operation
The instrument adjusts its full-scale sensitivity so that the magnitude output lies
between 30% and 90% of full-scale.
ASM Perform an Auto-Measure operation
The instrument adjusts its full-scale sensitivity so that the magnitude output lies
between 30% and 90% of full-scale, and then performs an auto-phase operation to

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 7230 and is the answer not in the manual?

Ametek 7230 Specifications

General IconGeneral
BrandAmetek
Model7230
CategoryAmplifier
LanguageEnglish

Related product manuals