M19-2101 REVISION 2.0 DVNEXT RHEOMETER - OPERATIONAL MANUAL
Measure the viscosity and record the Rheometer reading.
Note: The spindle must rotate at least five (5) times before readings are taken.
The viscosity reading should equal the cP value on the viscosity fluid standard to within the combined
accuracies of the rheometer and the standard (as discussed in the section entitled, Interpretation of
Calibration Test Results). However, instrument accuracy is ±2% of the maximum viscosity range and not
the standard 1%.
Calibration procedure for a Thermosel System
AMETEK Brookfield recommends a two-step check. First verify the calibration of the rheometer using the standard
rheometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles) as detailed in this appendix.
Second verify the calibration of the rheometer using the Thermosel. The use of an accessory device may decrease
the accuracy of measurement associated with the DVNext.
When a Thermosel System is used, the controller stabilizes the Thermo Container at the test temperature.
Put the proper amount of HT viscosity standard fluid into the HT-2 sample chamber. The amount varies
with the spindle used (refer to the Thermosel instruction manual).
Place the sample chamber into the Thermo Container.
Put the spindle into the test fluid and attach the extension link, coupling nut and free hanging spindle (or
directly attach the solid shaft spindle) to the DVNext.
Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test temperature.
Measure the viscosity and record the Rheometer reading. Note: The spindle must rotate at least five (5)
times before readings are taken.
The viscosity reading should equal the cP value on the viscosity fluid standard to within the combined
accuracies of the rheometer and the standard (as discussed in the section entitled, Interpretation of
Calibration Test Results). However, instrument accuracy is ±2% of the maximum viscosity range and not
the standard 1%.
Calibration procedure using UL or DIN adapters
AMETEK Brookfield recommends a two-step check. First, verify the calibration of the rheometer using the standard
rheometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles) as detailed in this appendix.
Second, verify the calibration of the rheometer using the UL or DIN Adapters. The use of an accessory device may
increase the accuracy of measurement associated with the DVNext.
When a UL or DIN Adapter is used, the water bath is stabilized at the proper temperature:
Put the proper amount of viscosity standard fluid into the UL Tube (refer to the UL Adapter instruction
manual).
Attach the spindle (with extension link and coupling nut) onto the DVNext.
Attach the tube to the mounting channel.
Lower the tube into the water bath reservoir, or if using the ULA-40Y water jacket, connect the inlet/
outlets to the bath external circulating pump.
Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test temperature.
Measure the viscosity and record the Rheometer reading.
Note: The spindle must rotate at least five (5) times before readings are taken.
The viscosity reading should equal the cP value on the viscosity fluid standard to within the combined
accuracies of the rheometer and the standard (as discussed in the section entitled, Interpretation of
Calibration Test Results). However, instrument accuracy is ±2% of the maximum viscosity range and not
the standard 1%.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek BROOKFIELD DVNext and is the answer not in the manual?

Ametek BROOKFIELD DVNext Specifications

General IconGeneral
BrandAmetek
ModelBROOKFIELD DVNext
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals