Dansensor® MAP Mix Provectus EN User Guide 45
P/N 300862-K
11/2019
Cleaning and Maintenance
General
Performing cleaning and maintenance regularly reduces the chances of equipment failure.
CAUTION! Personnel performing any maintenance or cleaning must familiarize
themselves with the "Safety Instructions" on page 9 before attempting any of these
procedures.
NOTE! For correct mounting/dismounting of fittings please see "Correct mounting/
dismounting of fittings" on page 17.
Spare parts
For a list of spare parts see "Spare parts, consumables and accessories" on page 75.
Cleaning
All of the device’s surfaces should only be cleaned using a mild soap solution and a wrung
cloth.
CAUTION! Never use hard tools or abrasive materials when cleaning any part of the
device.
WARNING! Never use cleaning agents containing chlorinated solvents or acetic or
phosphoric acid. These constitute a health hazard and could damage the
instrument.
Replacing gas inlet filters
Replace the gas inlet filters as described below:
Disconnect the gas inlet hose and the hose fitting from the gas inlet filter unit .
Remove the filter unit from the device.
Mount the new filter unit on the device using the gasket .
Connect the hose fitting using the gasket and then connect gas inlet hose .