Patient monitor User’s manual
9
automatically.
Monitor, Surgery mode: 50Hz/60Hz AC filter turns on
automatically.
Decreasing depth for AC filter: -20 dB
Electrode offset
potential
Monitor mode: ≤ 3 s; Surgery mode: ≤ 1 s
Indication of
electrode separation
Every electrode (exclusive of RL)
12.5 mm/s, 25 mm/s, 50 mm/s
Breakdown Voltage 4000VAC 50Hz/60Hz.
<5s after Defibrillation. (Mon or Surg mode)
Recovery time of
electrodes after
defibrillation
ECG waveform will recover to the baseline in 10 s.
With pacer pulse detector, complies with IEC
60601-2-27: 2005, 50.102.12. For the pacer pulse in
compliance with following conditions, pacer mark will
be signed on the screen(≥ 2 mm):
Pacing pulse amplitudes: ±2 mV to ±700mV
Pacing pulse widths: 0.1 ms to 2ms
Rise time: 10 us to 100us
Adult: 10 bpm to 300 bpm
Pediatric and Neonatal: 10 bpm to 350 bpm