Patient monitor User’s manual
9
line frequency)
automatically.
Monitor, Surgery mode: 50Hz/60Hz AC filter turns on
automatically.
Decreasing depth for AC filter: -20 dB
Input impedance
5.0 MΩ
ECG signal range
± 10.0 mV
Electrode offset
potential
± 500 mV
Patient leakage
current
< 10 uA
System noise
30 μVpp (RTI )
Standardizing signal
1 mV ± 5%
Baseline recovery
Monitor mode: 3 s; Surgery mode: 1 s
Indication of
electrode separation
Every electrode (exclusive of RL)
Sweep speed
12.5 mm/s, 25 mm/s, 50 mm/s
Protection
Breakdown Voltage 4000VAC 50Hz/60Hz.
Baseline recovery
<5s after Defibrillation. (Mon or Surg mode)
Recovery time of
electrodes after
defibrillation
ECG waveform will recover to the baseline in 10 s.
Pacer pulse mark
With pacer pulse detector, complies with IEC
60601-2-27: 2005, 50.102.12. For the pacer pulse in
compliance with following conditions, pacer mark will
be signed on the screen( 2 mm):
Pacing pulse amplitudes: ±2 mV to ±700mV
Pacing pulse widths: 0.1 ms to 2ms
Rise time: 10 us to 100us
HR
Measurement range
Adult: 10 bpm to 300 bpm
Pediatric and Neonatal: 10 bpm to 350 bpm
Resolution
1 bpm