Patient monitor User’s manual
10
Accuracy
±1% or ±1 bpm, whichever is greater
Detecting sensitivity
(II lead)
0.20mVpp
Response of HR to
QRS amplitude is
1mVp-p, width is
10ms
Adult: without response
Pediatric and Neonatal: with response
Alarm range
0 bpm to 350 bpm, high/low limit can be adjusted
continuously.
ST segment
Measurement
Channels
Calculating I, II, V- lead etc. at the same time
Default : II lead
Measurement range
-2.0 mV to +2.0 mV
Accuracy
-0.8 mV to +0.8 mV: ±0.02 mV or ±10%, whichever is
greater; Over ±0.8mV: unspecified
Resolution
0.01mV
Alarm range
-2.0 mV to +2.0 mV, high/low limit can be adjusted
continuously.
Arrhythmia analysis
Kinds
ASYSTOLE, VENT FIB/TACH, PAC, RUN PVCS,
COUPLET, BIGEMINY, TRIGEMINY, R on T,
TACHY, BRADY, MISSED BEATS, MULRI PACS,
PNP, PNC, NOISE, VPB, ST HIGH, ST LOW,
VTACH, PVCS HIGH
ECG/ST/arrhythmia supplemental information as required by AAMI
EC13
Electrosurgery
protection
electrocautery
uint protection
Cut mode: 300W
Coag mode: 100W
HR change: 10%
Resuming time: 10s
Complies with ANSI/AAMI EC 13:2002, 4.2.9.14
Input circuit current
< 0.1uA