Patient monitor User’s manual
13
Neonatal
Dia:
Adult
Pediatric
Neonatal
Map:
Adult
Pediatric
Neonatal
30 to 135
10 to 220
10 to 220
10 to 100
20 to 235
20 to 225
20 to 125
Pressure accuracy
Static:
Clinic:
±3 mmHg
Average error: ±5 mmHg, standard deviation: ≤8
mmHg
The cuff will deflate automatically when power is off
or time of measurement is beyond 120 seconds(90
seconds for neonate) or the cuff pressure is beyond the
overpressure protection set by software and hardware.
Normally, it is 20s to 45s (depending on HR and
moving interference typically)
Maximal measurement time: 120s( adult / pediatric),
90s(neonate)
Initial inflation
pressure
Adult default: 170 mmHg
Pediatric default: 130 mmHg
Neonatal default: 100 mmHg
Software
overpressure
protection
Adult: (297±3) mmHg
Pediatric: (252±3) mmHg
Neonatal: (147±3) mmHg
Assistant
venipuncture
inflation mode
Inflation pressure ( should be close to diastolic
pressure
)
:
Adult : 20 mmHg to 120 mmHg (normally 80 mmHg);